Free automatic software for quality assurance of computed tomography calibration, edges and radiomics metrics reproducibility.
Fecha de publicación:
Fecha Ahead of Print:
Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Saborido-Moral, Juan D.
- Cristian-Marin, Andrei
- Campayo-Esteban, Juan M.
- Baltas, Dimos
Grupos
Abstract
PURPOSE: To develop a QA procedure, easy to use, reproducible and based on open-source code, to automatically evaluate the stability of different metrics extracted from CT images: Hounsfield Unit (HU) calibration, edge characterization metrics (contrast and drop range) and radiomic features. METHODS: The QA protocol was based on electron density phantom imaging. Home-made open-source Python code was developed for the automatic computation of the metrics and their reproducibility analysis. The impact on reproducibility was evaluated for different radiation therapy protocols, and phantom positions within the field of view and systems, in terms of variability (Shapiro-Wilk test for 15 repeated measurements carried out over three days) and comparability (Bland-Altman analysis and Wilcoxon Rank Sum Test or Kendall Rank Correlation Coefficient). RESULTS: Regarding intrinsic variability, most metrics followed a normal distribution (88% of HU, 63% of edge parameters and 82% of radiomic features). Regarding comparability, HU and contrast were comparable in all conditions, and drop range only in the same CT scanner and phantom position. The percentages of comparable radiomic features independent of protocol, position and system were 59%, 78% and 54%, respectively. The non-significantly differences in HU calibration curves obtained for two different institutions (7%) translated in comparable Gamma Index G (1mm, 1%, >99%). CONCLUSIONS: An automated software to assess the reproducibility of different CT metrics was successfully created and validated. A QA routine proposal is suggested.
Copyright © 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 1120-1797, 1724-191X
- Tipo:
- Article
- Páginas:
- 103153-103153
- PubMed:
- 37778209
- Enlace a otro recurso:
- www.scopus.com
- Factor de Impacto:
- 0,747 SCImago ℠
- Cuartil:
- Q2 SCImago ℠
Physica Medica-European Journal of Medical Physics ELSEVIER SCI LTD
Citas Recibidas en Web of Science: 3
Citas Recibidas en Scopus: 3
Documentos
- No hay documentos
Filiaciones
Filiaciones no disponibles
Keywords
- Computed tomography; Automatic quality assurance; Radiomics; Reproducibility
Portal de investigación