Assaying multiple I-125 seeds with the well-ionization chamber SourceCheck(4 Pi) 33005 and a new insert

Autors de IIS La Fe
Autors aliens a IIS La Fe
- Ballester, F
- Vijande, J
Abstract
Purpose: To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck(4 Pi) 33005 (PTW, Germany) that allows evaluating the mean air-kerma strength of up to ten I-125 seeds with one single measurement instead of measuring each seed individually. Material and methods: The material required is: a) the SourceCheck(4 Pi) 33005 well-ionization chamber provided with a PTW insert to measure the air-kerma strength S-K of one single seed at a time; b) a newly designed insert that accommodates ten seeds in one column, which allows measuring the mean S-K of the ten seeds in one single measurement; and c) a container with ten seeds from the same batch and class of the seeds used for the patient implant, and a set of nine non-radioactive seeds.The new insert is characterized by determining its calibration coefficient, used to convert the reading of the well-chamber when ten seeds are measured to their mean S-K. The proposed method is validated by comparing the mean S-K of the ten seeds obtained from the new insert with the individual measurement of S-K of each seed, evaluated with the PTW insert. Results: The ratio between the calibration coefficient of the new insert and the calibration coefficient of the PTW insert for the SourceCheck(4 Pi) 33005 is 1.135 +/- 0.007 (k = 1). The mean S-K of a set of ten seeds evaluated with this new system is in agreement with the mean value obtained from measuring independently the S-K of each seed. Conclusions: The new insert and procedure allow evaluating the mean S-K of ten seeds prior to the implant in a single measurement. The method is faster and more efficient from radiation protection point of view than measuring the individual S-K of each seed.
Dades de la publicació
- ISSN/ISSNe:
- 1689-832X, 2081-2841
- Tipus:
- Article
- Pàgines:
- 492-496
- PubMed:
- 26816507
- Factor d'Impacte:
- 0,488 SCImago ℠
- Quartil:
- Q2 SCImago ℠
Journal of contemporary brachytherapy TERMEDIA PUBLISHING HOUSE LTD
Cites Rebudes en Web of Science: 2
Documents
- No hi ha documents
Filiacions
Keywords
- brachytherapy; insert; quality assurance; prostate; seeds; well chamber
Projectes associats
ESTUDIO DE BRAQUITERAPIA ELECTRÓNICA PARA CARCINOMA CUTÁNEO BASOCELULAR.
Investigador Principal: JOSÉ PÉREZ CALATAYUD
BCCBRACHY