Evaluation of the shielding in a treatment room with an electronic brachytherapy unit

Fecha de publicación:

Autores de IIS La Fe

  • Blanca Ibañez Rosello

    Autor

  • Juan Antonio Bautista Ballesteros

    Autor

  • Cristian Candela Juan

    Autor

  • Facundo Ballester Pallares

    Autor

  • Javier Vijande Asenjo

    Autor

  • José Pérez Calatayud

    Autor

Grupos

Abstract

Esteya (R) (Elekta Brachytherapy, Veenendaal, The Netherlands) is an electronic brachytherapy (eBT) system based on a 69.5 kVp x-ray source and a set of collimators of 1 to 3 cm in diameter, used for treating non-melanoma skin cancer lesions. This study aims to estimate room shielding requirements for this unit. The non-primary (scattered and leakage) ambient dose equivalent rates were measured with a Berthold LB-133 monitor (Berthold Technologies, Bad Wildbad, Germany). The latter ranges from 17 mSv h(-1) at 0.25 m distance from the x-ray source to 0.1 mSv h(-1) at 2.5 m. The necessary room shielding was then estimated following US and some European guidelines. The room shielding for all barriers considered was below 2 mmPb. The dose to a companion who, exceptionally, would stay with the patient during all treatment was estimated to be below 1 mSv if a leaded apron is used. In conclusion, Esteya shielding requirements are minimal.

Datos de la publicación

ISSN/ISSNe:
0952-4746, 1361-6498

JOURNAL OF RADIOLOGICAL PROTECTION  IOP Publishing Ltd

Tipo:
Article
Páginas:
5-12
PubMed:
28253202
Factor de Impacto:
0,848 SCImago
Cuartil:
Q2 SCImago

Citas Recibidas en Web of Science: 5

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Esteya; electronic brachytherapy; shielding; radiation protection

Cita

Compartir