Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer

Fecha de publicación:

Autores de IIS La Fe

Participantes ajenos a IIS La Fe

  • Pardo, A

Abstract

BackgroundAccurate evaluation of hematology analyzers is recommended before these devices can be broadly introduced for the routine testing of continuous ambulatory peritoneal dialysis (CAPD), ascitic, and pleural fluids. MethodsWe evaluated the performance of Mindray BC-6800 for white blood cell (WBC) and differential cell count in 50 CAPD, 60 ascitic and 40 pleural compared with manual microscopy. Within-run precision, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), and carryover were assessed. ResultsThe Passing-Bablok regression in all fluids showed the following equations: y(WBC)=1.05x+3.31 (95%CI slope 0.95 to 1.12; intercept -0.25 to 5.52); y(MN)=0.85x+15.63 (95%CI slope 0.72 to 1.05; intercept -24.18 to 84.47); and y(PMN)=1.21x+13.37 (95%CI slope 1.03 to 1.35; intercept 4.00 to 32.47) with bias 78 cells/L. The AUC for clinical PMN cut-off was 0.88 (95%CI: 0.77 to 0.98). In ascitic, pleural, and CAPD fluids the AUC for clinical PMN cut-off were 0.88 (95%CI: 0.63 to 1.00), 0.83 (95%CI: 0.68 to 0.99), and 1.00 (95%CI: 1.00 to 1.00) respectively. CV ranged from 3%-34%. LoB of 3cell/L was verified. LoD and LoQ reported the same result (8cells/L). Carry over never exceeded 0.05%. ConclusionThe effectiveness of BC-6800 to categorize cells from different body fluids was not compromised by the slight positive bias observed. This conclusion is supported by the high AUC and agreement between the automated method and the reference method. The results show that BC-6800 offers rapid, accurate, and reproducible results for clinical management of CAPD, ascitic, and pleural fluids.

Datos de la publicación

ISSN/ISSNe:
0887-8013, 1098-2825

JOURNAL OF CLINICAL LABORATORY ANALYSIS  WILEY

Tipo:
Article
Páginas:
-
PubMed:
28444939
Factor de Impacto:
0,509 SCImago
Cuartil:
Q2 SCImago

Citas Recibidas en Web of Science: 8

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • automated hematology analyzer; cell count; differential count; serous fluid

Compartir