Procedure for automated background correction in flow systems with infrared spectroscopic detection and changing liquid-phase composition.

Fecha de publicación:

Autores de IIS La Fe

Participantes ajenos a IIS La Fe

  • Lendl B

Abstract

This paper describes a partial least squares (PLS) based automatic procedure to correct for changes in the spectral contribution of the solvent or solvent mixtures from solute spectra recorded in these solvents. The procedure was developed for successful on-line Fourier transform infrared (FT-IR) detection in gradient high-performance liquid chromatography (HPLC) separations. It requires a reference FT-IR data set containing all possible combinations of the expected variation in solvent composition. Furthermore, a spectral region (A) in these spectra is required where the solvents show absorption but the analytes do not. This is the case for the system ACN:H(2)O, an often-applied solvent mixture in gradient HPLC, in the spectral region from 2300-2400 cm(-1). By focusing on (A) the developed numerical method selects an appropriate background spectrum from the reference data set, which is then subtracted from the analyte spectra. The method was programmed in Matlab and tested on different isocratic and gradient on-line reversed phase liquid chromatography-Fourier transform infrared spectrometry (LC-FT-IR) data sets. This work describes a successful method to perform eluent correction in on-line coupling of FT-IR spectrometry with gradient LC.

Datos de la publicación

ISSN/ISSNe:
0003-7028, 1943-3530

APPLIED SPECTROSCOPY  SAGE PUBLICATIONS INC

Tipo:
Article
Páginas:
1363-1369
PubMed:
20030981
Factor de Impacto:
0,635 SCImago
Cuartil:
Q2 SCImago

Citas Recibidas en Web of Science: 8

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Cita

Compartir