Evaluation of the efficacy of the automatic ozone decontamination device in a real-life setting of an intravitreal injection room.

Fecha de publicación: Fecha Ahead of Print:

Autores de IIS La Fe

  • Mª Mercedes Salvador Aguilá

    Autor

  • Clara Martinez Rubio

    Autor

  • Agnieszka Janicka Caulineau

    Autor

  • Marta Moro Muñiz

    Autor

Participantes ajenos a IIS La Fe

  • Adloff-Alonso A.
  • Rodrigo R.

Grupos

Abstract

OBJECTIVE: The objective of this study was to evaluate the efficacy of an ozonation system for disinfecting intravitreal injection rooms. DESIGN: A prospective intervention pilot study was conducted with a 'before-and-after' design. SETTING: The study was conducted in an intravitreal injection room at Hospital de Manises. METHODS: Air and surface samples were taken before and after ozonation using a remote-controlled ozonation system. The room was treated with ozone at a concentration of 9-10 ppm for 20 min. Measurements were taken immediately after, 12 h, and 24 h post ozonation. Statistical analysis was performed using the Wilcoxon signed-rank test. RESULTS: At 24 h post ozonation, significant reductions in microbial load were observed: in the centre of the room, the bacterial load reduced from >250 colony-forming units (cfu) to 10 cfu (96%) and fungi from 14 cfu to <1 cfu (93.1%) (P < 0.05); in the diffuser, the bacterial load reduced from 61 cfu to 24 cfu (60.7%) and fungi from 6 cfu to <1 cfu (100%) (P < 0.05) and on the procedure chair, the bacterial load reduced from 51 cfu to 7 cfu (86.3%) and fungi from 6 cfu to 3 cfu (50%) (P < 0.05). CONCLUSIONS: In a dedicated intravitreal injection room, a remotely controlled ozonation cycle at 9-10 ppm for 20 min was associated with substantial short-term reductions in airborne and selected surface microbial load under unoccupied conditions. Automated ozonation may complement routine cleaning and ventilation to maintain low environmental bioburden; its clinical impact remains to be determined.

© 2026 The Authors.

Datos de la publicación

ISSN/ISSNe:
2590-0889, 2590-0889

Infection Prevention in Practice  ELSEVIER

Tipo:
Article
Páginas:
100507-100507
PubMed:
41684792
Enlace a otro recurso:
www.scopus.com

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Automatic room disinfection; Intravitreal injection; Ozone

Compartir