Acute hemodynamic effects of a novel algorithm for cardiac resynchronization therapy optimization: Results from the BIO|Adapt study
Fecha de publicación:
Fecha Ahead of Print:
Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Garcia-Fernandez F.J.
- Ando K.
- Kato R.
- Martínez J.G.
- Berruezo A.
- Jimenez J.
- Anneken L.
- Hironobe N.
- Harada M.
- Yagishita A.
- Mitkowski P.
- Hayashi H.
- Kishihara J.
- Pardo-Fresno M.
- Ebrahim I.
- Meyhöfer J.
- Salguero-Bodes R.
- Marques P.
- Moreno A.
- Pinart M.
- Bulava A.
Grupos
Abstract
Background A considerable proportion of patients with heart failure with reduced left ventricular ejection fraction (LVEF) and prolonged QRS duration fail to respond adequately to cardiac resynchronization therapy (CRT). Suboptimal atrioventricular (AV) delay programming is a key factor leading to diminished response. Objective This study aimed to acutely correlate aortic velocity time integral, LVEF, and QRS duration with 3 programming strategies in the postimplantation setting: AV delay automatically set by a novel CRT AutoAdapt algorithm, AV delay optimized by echocardiography, and standard (“in-box”) programming. Methods Under normal intrinsic AV conduction (<250 ms after atrial pacing), the CRT AutoAdapt algorithm automatically switches to left ventricular-only pacing and adapts the AV delay to 50%, 70% (default), or 90% of intrinsic AV conduction. In our study, 163 CRT-defibrillator recipients (mean age 69.2 ± 9.2 years; 32% female; LVEF 27.7% ± 6.6%; 76% with left bundle branch block) underwent echocardiographic and electrocardiographic assessments at prehospital discharge. Results With the default 70% AV adaptation, the CRT AutoAdapt automatically set a similar AV delay (117.1 ± 23.9 ms) compared with echocardiography-guided optimization (118.3 ± 30.2 ms; P = .94). Strong concordance was observed between the algorithm and echocardiography-guided optimization for aortic velocity time integral (concordance correlation coefficient 0.96) and LVEF (0.93). Both optimization methods significantly outperformed standard AV programming for LVEF improvement ( P < .001) and QRS-width reduction ( P < .01), with no significant differences between optimization approaches. Conclusion Acute hemodynamic effects of CRT AutoAdapt and echocardiography-guided AV-delay optimization were similar, with equivalent cardiac performance. This device-based approach offers the potential advantage of continuous adaptation to changing physiological conditions without requiring resource-intensive in-office optimization procedures. © 2026 Heart Rhythm Society.
Datos de la publicación
- ISSN/ISSNe:
- 2666-5018, 2666-5018
- Tipo:
- Article
- Páginas:
- 686-694
- PubMed:
- 42038676
- Enlace a otro recurso:
- www.scopus.com
Heart Rhythm O2 ELSEVIER
Documentos
- No hay documentos
Filiaciones
Keywords
- beta adrenergic receptor blocking agent; dipeptidyl carboxypeptidase inhibitor; aged; algorithm; Article; atrioventricular conduction; cardiac resynchronization therapy; clinical trial; coronary artery bypass graft; diabetes mellitus; electrocardiography; female; good clinical practice; heart left ventricle ejection fraction; hemodynamics; human; hypertension; male; patient selection; post hoc analysis; prospective study; QRS interval; sensitivity analysis; valvular heart disease
Portal de investigación