Comparison of Cone-Beam CT Angiography and Contrast-Enhanced CT Guidance Using Electromagnetic Navigation for Percutaneous Liver Microwave Ablation: A Retrospective Nonrandomized Observational Study
Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Lloret, Maximiliano R.
- Pisoni, Serena
- Arminana, Arturo
- Mora, Daniel F.
- Picado, Adrian
- Martinez, Jose J.
- Alcaide, Julien
Grupos
Abstract
Purpose: To evaluate whether C-arm cone-beam computed tomography (CT) angiography guidance is noninferior to standard contrast-enhanced CT guidance for hepatic microwave ablation (MWA) using an electromagnetic navigation system (Imactis CT-Navigation System) in tumors not visible on ultrasound (US). Materials and Methods: This retrospective, single-center, nonrandomized study included 79 patients with liver neoplasms treated with MWA between September 2019 and June 2024. Patients were allocated to cone-beam CT angiography (n = 53) or contrast-enhanced CT (n = 26) groups based on modality availability. The primary outcome was technique effectiveness, defined by absence of tumor on imaging immediately postprocedurally, at 1-3 months, and at 4-6 months. Secondary end points included procedural time, effective radiation dose, and 30-day adverse events. Comparisons used t Results: Technique effectiveness was comparable between cone-beam CT angiography and contrast-enhanced CT at all time points: immediate (technical success), 100% vs 100% (P = 1.000); 1-3 months, 87% vs 73% (P = .208); and 4-6 months, 81% vs 71% (P = .463). No significant differences were found in radiation dose (37.96 vs 32.36 mSv; P = .276) or adverse event rates (11% vs 11%, P = 1.000). Two deaths occurred in the cone-beam CT angiography group: one due to hospital-acquired pneumonia and the other from colonic perforation, possibly related to inadequate technique selection. Conclusions: Cone-beam CT angiography-guided MWA using electromagnetic navigation offered comparable short-term effectiveness and safety to contrast-enhanced CT guidance, but with longer procedural times. This small nonrandomized cohort study was subject to selection bias, and further studies are needed to confirm these findings.
Copyright © 2025 SIR. Published by Elsevier Inc. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 1051-0443, 1535-7732
- Tipo:
- Article
- Páginas:
- 1696-1704
- PubMed:
- 40754299
- Factor de Impacto:
- 0,914 SCImago ℠
- Cuartil:
- Q1 SCImago ℠
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY ELSEVIER SCIENCE INC