Safety and effectiveness of the first balloon-in-basket pulsed field ablation system for the treatment of atrial fibrillation: VOLT CE Mark Study 6-month results.
Fecha de publicación:
Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Tilz RR
- Chierchia GB
- Gunawardene M
- Sanders P
- Haqqani H
- Kalman J
- Healy S
- Pürerfellner H
- Neuzil P
- Loh P
- Reddy VY
- Knecht S
- Jesser E
- Dirckx N
- Miller A
- Walker D
- Lakkireddy D
Grupos
Abstract
AIMS: Pulsed field ablation (PFA) is a growing ablation modality for pulmonary vein isolation (PVI) in atrial fibrillation (AF) patients. This study assesses the 6-month safety and effectiveness of a novel balloon-in-basket, mapping-integrated PFA system, with a purpose-built form factor for PVI. METHODS AND RESULTS: The VOLT CE Mark Study is a prospective, multi-center, pre-market study. A total of 150 patients with drug-refractory paroxysmal (PAF) or persistent AF (PersAF) were enrolled between 8 November 2023 and 14 March 2024, of which 146 patients (age 64.1 ± 10.0 years, 63.0% male, 70.5% PAF) underwent PVI with the balloon-in-basket PFA catheter and system featuring integrated electroanatomic mapping with contact-sensing. Study endpoints were the rate of primary serious adverse events within 7 days as well as acute procedural effectiveness and 6-month freedom from recurrence. Acute effectiveness was achieved in 99.1% (573/578) of treated PVs (98.6% of patients, 144/146) with 17.6 ± 5.7 PFA applications/patient. Procedure, fluoroscopy, LA dwell, and transpired ablation times were 100.4 ± 33.0, 17.3 ± 12.1, 39.4 ± 20.6, and 31.4 ± 16.8 min, respectively. There were 4 (2.7%; 4/146) primary serious adverse events. The rate of freedom from documented atrial arrhythmias was 88.2% in PAF patients and 76.7% in PersAF patients (freedom from symptomatic recurrence was documented in 90.2% of PAF patients and 74.4% of PersAF patients) through 6-months post-index procedure. CONCLUSION: The VOLT CE Mark Study primary results demonstrate the safety and effectiveness of the novel balloon-in-basket PFA system to perform PVI in PAF and PersAF.
© The Author(s) 2025. Published by Oxford University Press on behalf of the European Society of Cardiology.
Datos de la publicación
- ISSN/ISSNe:
- 1099-5129, 1532-2092
- Tipo:
- Article
- Páginas:
- -
- PubMed:
- 40163671
- Enlace a otro recurso:
- www.scopus.com
- Factor de Impacto:
- 1,957 SCImago ℠
- Cuartil:
- Q1 SCImago ℠
EUROPACE OXFORD UNIV PRESS
Citas Recibidas en Web of Science: 59
Documentos
- No hay documentos
Filiaciones
Keywords
- Atrial fibrillation; Balloon; Balloon-in-basket; Basket; Catheter ablation; Pulmonary vein isolation; Pulsed field ablation; Single-shot
Cita
Tilz RR,Chierchia GB,Gunawardene M,Sanders P,Haqqani H,Kalman J,Healy S,Pürerfellner H,Neuzil P,ASENSI JO,Loh P,Reddy VY,Knecht S,Jesser E,Dirckx N,Miller A,Walker D,Lakkireddy D. Safety and effectiveness of the first balloon-in-basket pulsed field ablation system for the treatment of atrial fibrillation: VOLT CE Mark Study 6-month results. Europace. 2025. 27. (4):euaf072. IF:10,200. (1).
Portal de investigación