Image reject analysis and rejection causes in digital radiography at a university hospital through estimates of patient doses.

Fecha de publicación: Fecha Ahead of Print:

Autores de IIS La Fe

Grupos

Abstract

The introduction of digital radiography has improved image acquisition. However, rejection of images remains a matter of concern. Reject analysis is part of the quality assurance program in radiology and helps identify potential errors or lack of training. A retrospective study was conducted at the radiology department of a university hospital. The reject rate was calculated both using the number of examinations, $r_n$, and the dose-area product, $r_d$. A reject rate $r_n$ of 3.3% for paediatric units and 4.5% for adults was found. The corresponding values of rd were 4.4 and 8.4%, respectively. The main rejection cause was patient motion, being 50.2% of rejected examinations in adults and 63.7% in children. The contribution of exposure errors was minor, as expected in digital radiography units. A discrepancy between reject rates $r_n$ and $r_d$ was observed, suggesting dosimetric quantities could be considered in reject analysis for further assessment of patient radiation burden.

© The Author(s) 2023. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Datos de la publicación

ISSN/ISSNe:
0144-8420, 1742-3406

RADIATION PROTECTION DOSIMETRY  OXFORD UNIV PRESS

Tipo:
Article
Páginas:
274-284
Factor de Impacto:
0,355 SCImago
Cuartil:
Q3 SCImago

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • RATES; FILM

Proyectos asociados

REALIZING THE EUROPEAN NETWORK IN BIODOSIMETRY (RENEB)

295513_RENEB_PE_MONTORO_FP7-Fission-2011 . FUNDACIÓN PARA LA INVESTIGACIÓN DEL HOSPITAL UNIVERSITARIO LA FE DE LA COMUNIDAD VALENCIANA . 2012

CONTROL Y SEGUIMIENTO INTEGRAL DE LA RADIACIÓN IONIZANTE RECIBIDA TRAS EXPLORACIONES MÉDICAS: COSIRI.

Investigador Principal: ÁNGEL ALBERICH BAYARRI

GV/2015/132 . FUNDACIÓN PARA LA INVESTIGACIÓN DEL HOSPITAL UNIVERSITARIO LA FE DE LA COMUNIDAD VALENCIANA; CONSELLERIA DE EDUCACION . 2015

Papel de la biodosimetría en la optimización del uso de 131I-MIBG como intensificación del tratamiento del neuroblastoma de alto riesgo.

Investigador Principal: JULIA BALAGUER GUILL

JBG-IOD-2017-01 . 2024

Estudio comparativo de ablación de tumores hepáticos empleando sistema de navegación IMACTIS mediante TIC y CBCT.

Investigador Principal: MAXIMILIANO LLORET LARREA

IMA-CT-CBCT . 2023

Cita

Compartir