Incidence of premature battery depletion in subcutaneous cardioverter-defibrillator patients: insights from a multicenter registry.

Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Luker, Jakob
- Strik, Marc
- Andrade, Jason G
- Raymond-Paquin, Alexandre
- Elrefai, Mohamed Hassan
- Roberts, Paul R
- Kron, Jordana
- Koneru, Jayanthi
- Franqui-Rivera, Hilton
- Sultan, Arian
- Ernst, Angela
- Schmitt, Jorn
- Pott, Alexander
- Veltmann, Christian
- Srinivasan, Neil T
- Collinson, Jason
- van Stipdonk AMW
- Linz, Dominik
- Fluschnik, Nina
- Tönnis T
- Haeberlin, Andreas
- Ploux, Sylvain
- Steven, Daniel
Grupos
Abstract
BACKGROUND: The subcutaneous ICD established its role in the prevention of sudden cardiac death in recent years. The occurrence of premature battery depletion in a large subset of potentially affected devices has been a cause of concern. The incidence of premature battery depletion has not been studied systematically beyond manufacturer-reported data. METHODS: Retrospective data and the most recent follow-up data on S-ICD devices from fourteen centers in Europe, the US, and Canada was studied. The incidence of generator removal or failure was reported to investigate the incidence of premature S-ICD battery depletion, defined as battery failure within 60 months or less. RESULTS: Data from 1054 devices was analyzed. Premature battery depletion occurred in 3.5% of potentially affected devices over an observation period of 49 months. CONCLUSIONS: The incidence of premature battery depletion of S-ICD potentially affected by a battery advisory was around 3.5% after 4 years in this study. Premature depletion occurred exclusively in devices under advisory. This is in line with the most recently published reports from the manufacturer. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT04767516 .
© 2023. The Author(s).
Datos de la publicación
- ISSN/ISSNe:
- 1383-875X, 1572-8595
- Tipo:
- Article
- Páginas:
- -
- PubMed:
- 36652082
- Factor de Impacto:
- 0,680 SCImago ℠
- Cuartil:
- Q2 SCImago ℠
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY Kluwer Academic Publishers
Citas Recibidas en Web of Science: 1
Documentos
- No hay documentos
Filiaciones
Keywords
- Advisory; Implantable cardioverter-defibrillator; Longevity; Premature battery depletion; Subcutaneous ICD
Proyectos y Estudios Clínicos
INCORPORACIÓN GRUPOS CIBER. DR. DOLZ
Investigador Principal: LUIS VICENTE MARTÍNEZ DOLZ
CB16/11/00261 . INSTITUTO DE SALUD CARLOS III . 2017
Validación prospectiva de un índice pronóstico de muerte súbita o arritmias ventriculares malignas en adultos con cardiopatías congénitas.
Investigador Principal: TOMAS DATINO ROMANIEGA
PI17/01327 . 2018
MORE-CRT: EL AUMENTO DE LAS OPCIONES DISPONIBLES CON UN ELECTRODO VI TETRAPOLAR OFRECE SOLUCIONES CLÍNICAS A LOS PROBLEMAS DE LA TRC. MORE OPTIONS AVAILABLE WITH A QUADRIPOLAR LV LEAD PROVIDE IN CLINIC SOLUTIONS TO CRT CHALLENGES.
Investigador Principal: JOAQUÍN OSCA ASENSI
CR-11-006-HF-ID . 2012
GLORIA - AF: REGISTRO GLOBAL DEL TRATAMIENTO ANTITROMBÓTICO ORAL A LARGO PLAZO EN PACIENTES CON FIBRILACIÓN AURICULAR (FASE II/III – ESTADOS MIEMBROS DE LA UE/ AEE).
Investigador Principal: JOAQUÍN OSCA ASENSI
BOE-DAB-2012-01 . 2013
MORE RESPONSE ON CARDIAC RESYNCHRONIZATION THERAPY (CRT) WITH MULTIPOINT PACING (MPP) – MAYOR RESPUESTA A LA TERAPIA DE RESINCRONIZACIÓN CARDIACA (TRC) CON ESTIMULACIÓN MULTISITIO.
Investigador Principal: JOAQUÍN OSCA ASENSI
CR-13-006-ID-HF . 2013
ESTIMULACIÓN TRICAMERAL EN PACIENTES CON CARDIOMIOPATÍA HIPERTRÓFICA OBSTRUCTIVA. TRICHAMPION (TRIPLE CHAMBER PACING IN HOCM PATIENTS).
Investigador Principal: MARÍA JOSÉ SANCHO-TELLO DE CARRANZA
CR-11-030-EU-HF . 2014
ESTIMULACIÓN BIVENTRICULAR TRIFOCAL PARA LA OPTIMIZACIÓN DE LA TERAPIA DE RESINCRONIZACIÓN CARDIACA (TRC).
Investigador Principal: ÓSCAR CANO PÉREZ
ICSY02 . 2015
EVALUACIÓN CLÍNICA DEL PACIENTE Y FUNCIONAL DEL DISPOSITIVO EN LA PRÁCTICA CLÍNICA HABITUAL CON PORTADORES DE DAI.
Investigador Principal: MARÍA JOSÉ SANCHO-TELLO DE CARRANZA
RTSY03 . 2015
OPCIONES DE LA TERAPIA DE RESINCRONIZACIÓN CON EL ELECTRODO QUARTET TM (QUARTET LEAD AND RESYNCHRONIZATION THERAPY OPTIONS) III.
Investigador Principal: JOAQUÍN OSCA ASENSI
CRD_789 . 2015
REGISTRO DE ESTIMULACIÓN TRANSCATÉTER CON EL SISTEMA MICRATM (MICRA)
Investigador Principal: JOAQUÍN OSCA ASENSI
ESTUDIO MICRA PM . 2017
ESTIMULACIÓN SÍNCRONA MULTIPUNTO FRENTE A ESTIMULACIÓN SECUENCIAL MULTIPUNTO DEL VENTRÍCULO IZQUIERDO PARA TRC. ESTUDIO SYNSEQ.
Investigador Principal: JOAQUÍN OSCA ASENSI
ESTUDIO SYNSEQ . 2016
APNEA DE SUEÑO, ARRITMIAS Y REMODELADO CARDIACO INVERSO EN PACIENTES CON INSUFICIENCIA CARDIACA.
Investigador Principal: MARÍA JOSÉ SANCHO-TELLO DE CARRANZA
ICSY03 . 2017
EVALUACIÓN CLÍNICA DEL ALGORITMO DE AUTOUMBRAL DE VENTRÍCULO IZQUIERDO.
Investigador Principal: ÓSCAR CANO PÉREZ
IPTL01 . 2017
CONTROL ESTRATÉGICO PARA LA OPTIMIZACIÓN DE LA RESPUESTA A LA TERAPIA DE RESINCRONIZACIÓN CARDÍACA (STRATEGIC MANAGEMENT TO OPTIMIZE RESPONSE TO CARDIAC RESYNCHRONIZATION THERAPY) (SMART-CRT STUDY).
Investigador Principal: MARÍA JOSÉ SANCHO-TELLO DE CARRANZA
C2067 . 2017
ANTICOAGULANTES ORALES NO ANTAGONISTAS DE LA VITAMINA K EN PACIENTES CON EPISODIOS DE FRECUENCIA AURICULAR RÁPIDA.
Investigador Principal: JOAQUÍN OSCA ASENSI
NOAH-AFNET6 . 2018
Prevención de terapias inapropiadas y eventos adversos cardiacos y cerebrovasculares graves con un Desfibrilador Automático Implantable monocameral con detección auricular con dipolo flotante DX.
Investigador Principal: JOAQUÍN OSCA ASENSI
TRI-DX-2018-01 . 2018
EVALUACIÓN DE SEGURIDAD Y COMPORTAMIENTO ELÉCTRICO DE LAS SONDAS INVICTA EQUIPADAS CON CONECTOR DF4.
Investigador Principal: MARÍA TERESA IZQUIERDO DE FRANCISCO
LCN001 . 2019
ESTUDIO OASIS-H. ORGANIZACION ASISTENCIAL DE PACIENTES CON HOME MONITORING.
Investigador Principal: MARÍA JOSÉ SANCHO-TELLO DE CARRANZA
OASIS-HM . 2010
ESTUDIO ABIERTO, FASE IV, MULTICÉNTRICO, PARA EVALUAR LA EFICACIA DIAGNÓSTICA DE NUUBO NECGBAND ANTES DEL HOLTER IMPLANTABLE.
Investigador Principal: ÓSCAR CANO PÉREZ
SYN201401 . 2015
EVALUACION CLINICA DE UNA NUEVA FAMILIA DE ELECTRODOS DE FIJACION ACTIVA DE SORIN GROUP.
Investigador Principal: JOAQUÍN OSCA ASENSI
LBFX04
ENSAYO DE RESULTADOS PARA EVALUAR LA EFICACIA Y LA SEGURIDAD DE NORDITROPIN® EN PACIENTES ADULTOS QUE RECIBEN HEMODIALISIS CRONICA. ENSAYO ALEATORIZADO, DOBLE CIEGO, DE GRUPOS PARALELOS, CONTROLADO CON PLACEBO Y MULTICENTRICO.
Investigador Principal: Pablo Alcocer Yuste
NN-1630-1453
INFLUENCIA DE LOS DATOS DIAGNÓSTICOS PARA EL TRATAMIENTO DE LA ENFERMEDAD Y RELACIÓN DE LOS POLIMORFISMOS GENÉTICOS CON LA TAQUIARRITMIA VENTRICULAR EN PACIENTES PORTADORES DE DAI.
Investigador Principal: JOAQUÍN OSCA ASENSI
DISCOVERY
COMPARACION DE LA ESTIMULACION EN SEPTO DE VENTRICULO DERECHO VERSUS ESTIMULACION APICAL EN PACIENTES PORTADORES DE UN DESFIBRILADOR IMPLANTABLE CON TERAPIA DE RESINCRONIZACION CARDIACA.
Investigador Principal: JOAQUÍN OSCA ASENSI
SEPTALCRT04-08
ESTUDIO DE BUSQUEDA DE DOSIS, DOBLE CIEGO Y CONTROLADO CON PLACEBO SOBRE LA EFICACIA Y LA SEGURIDAD DE 50, 100 O 300 MG DIARIOS DE CELIVARONA, CON AMIODARONA COMO FARMACO DE REFERENCIA, PARA PREVENIR LAS INTERVENCIONES DEL DESFIBRILADOR CARDIOVERSOR IMPLA NTABLE O LA MUERTE.
Investigador Principal: JOSE OLAGUE DE ROS
DRI10936
ESTUDIO ESPAÑOL DE FIBRILACION AURICULAR Y RESINCRONIZACION
Investigador Principal: JOAQUÍN OSCA ASENSI
SPAREIII
PLAN DE INVESTIGACION CLINICA PARA EL ESTUDIO MASTER STUDY DE LINOXSMART S DX.
Investigador Principal: JOSE OLAGUE DE ROS
LINOXDXPME/MASTER
QUARTET LEAD AND RESYNCHRONIZATION THERAPY OPTIONS.
Investigador Principal: JOSE OLAGUE DE ROS
CR-10-046-SP-HF
ESTUDIO DE LA EFICACIA Y LA SEGURIDAD DE F373280 EN EL MANTENIMIENTO DEL RITMO SINUSAL TRAS CARDIOVERSIÓN ELÉCTRICA EN PACIENTES CON FIBRILACIÓN AURICULAR PERSISTENTE E INSUFICIENCIA CARDÍACA CRÓNICA. INTERNACIONAL, MULTICÉNTRICO, ALEATORIZADO, DOBLE CIEG O Y CONTROLADO CON PLACEBO.
Investigador Principal: JOAQUÍN OSCA ASENSI
F373280CA201
ESTUDIO OBSERVACIONAL PARA LA EVALUACIÓN DEL MANEJO DE PACIENTES RESINCRONIZADOS CON EL ELECTRODO TETRAPOLAR QUARTETTM DE VENTRÍCULO IZQUIERDO.
Investigador Principal: JOSE OLAGUE DE ROS
CR-12-015-SP-HF
REDUCCIÓN DE CHOQUES INAPROPIADOS CON EL DISCRIMINADOR DE RITMO PARAD+.
Investigador Principal: JOSE OLAGUE DE ROS
ITSY09
EVALUACIÓN NO INVASIVA DE LA ESTIMACIÓN DEL RIESGO POST-INFARTO - EFICACIA DE UN DAI.
Investigador Principal: JOAQUÍN OSCA ASENSI
REFINE_ICD
SEGURIDAD Y RENDIMIENTO DE LAS SONDAS DF4 VENTRICULARES INVICTA CON FIJACIÓN ACTIVA.
Investigador Principal: MARÍA TERESA IZQUIERDO DE FRANCISCO
LINI01 . 2020
APLICACIONES DE RADIOFRECUENCIA DE ALTA POTENCIA PARA ABLACIÓN DE VENAS PULMONARES MÁS RÁPIDA Y SEGURA.
Investigador Principal: JOAQUÍN OSCA ASENSI
POWERFAST II . 2020
OBSERVATION OF CLINICAL ROUTINE CARE FOR HEART FAILURE PATIENTS IMPLANTED WITH BIOTRONIK CRT DEVICES.
Investigador Principal: MARÍA JOSÉ SANCHO-TELLO DE CARRANZA
CR021 / CR028 . 2020
LEAD EVALUATION FOR DEFIBRILLATION AND RELIABILITY (LEADR).
Investigador Principal: ÓSCAR CANO PÉREZ
MDT19004 . 2021
EVALUACIÓN DE LA MICRO SONDA TETRAPOLAR AXONE PARA LA TERAPIA DE RESINCRONIZACIÓN CARDÍACA MEJORADA.
Investigador Principal: JOAQUÍN OSCA ASENSI
LAXI01 . 2021
ESTIMULACIÓN DEL SISTEMA DE CONDUCCIÓN Y ABLACIÓN DEL NODO AURÍCULOVENTRICULAR EN PACIENTES CON INSUFICIENCIA CARDIACA, FRACCIÓN DE EYECCIÓN DE VENTRÍCULO IZQUIERDO >40% Y FIBRILACIÓN AURICULAR PERMANENTE: el estudio PACE-FI
Investigador Principal: ÓSCAR CANO PÉREZ
PACE-FIB . 2021
Estudio aleatorizado de no inferioridad comparando la estimulación del área de la rama izquierda con la estimulación biventricular en pacientes con indicación de terapia de resincronización cardiaca: Estudio LEFT-BUNDLE-CRT.
Investigador Principal: ÓSCAR CANO PÉREZ
LB-CRT-001 . 2022
ABLACIÓN DE SUSTRATO COMO TRATAMIENTO CONCOMITANTE EN LA TAQUICARDIA AURICULAR MACROREENTRANTE DE AURÍCULA IZQUIERDA”
Investigador Principal: JOAQUÍN OSCA ASENSI
SLICE LAMR . 2022
Registro SEC – EXCELENTE FA: una herramienta para la mejora de la calidad de la atención médica a pacientes con fibrilación auricular en España.
Investigador Principal: JOAQUÍN OSCA ASENSI
SEC-FA-2022-1 . 2022
Nueva herramienta no invasiva para la valoración de los mecanismos electrofisiológicos en arritmias cardiacas. (SAVE-COR).
Investigador Principal: JOAQUÍN OSCA ASENSI
SAVE-COR . 2022