Validación preclínica de un respirador de turbina para la ventilación invasiva: el respirador ACUTE-19.

Fecha de publicación: Fecha Ahead of Print:

Autores de IIS La Fe

Participantes ajenos a IIS La Fe

  • Casan-Pallardo, M.
  • Redondo-Garcia, I, J.
  • Viscasillas-Monteagudo, J.
  • Gutierrez-Bautista, A.
  • Ramirez-Faz, J.
  • Alonso-Perez, P.
  • Diaz-Lobato, S.
  • Neto, A. S.
  • Gama de Abreu M
  • Pelosi, P.
  • Schultz, M. J.
  • open source smart breathing system ACUTE-19 investigators

Grupos

Abstract

BACKGROUND: The severe acute respiratory syndrome-coronavirus 2 pandemic pressure on healthcare systems can exhaust ventilator resources, especially where resources are restricted. Our objective was a rapid preclinical evaluation of a newly developed turbine-based ventilator, named the ACUTE-19, for invasive ventilation. METHODS: Validation consisted of (a) testing tidal volume delivery in 11 simulated models, with various resistances and compliances; (b) comparison with a commercial ventilator (VIVO-50) adapting the United Kingdom Medicines and Healthcare products Regulatory Agency-recommendations for rapidly manufactured ventilators; and (c) in vivo testing in a sheep before and after inducing acute respiratory distress syndrome by saline lavage. RESULTS: Differences in tidal volume in the simulated models were marginally different (largest difference 33 ml [95% CI 31 to 36]; P < .001). Plateau pressure was not different (-0.3 cmH(2)O [95% CI -0.9 to 0.3]; P = .409), and positive end-expiratory pressure was marginally different (0.3 cmH(2)O [95% CI 0.2 to 0.3]; P < .001) between the ACUTE-19 and the commercial ventilator. Bland-Altman analyses showed good agreement (mean bias -0.29 [limits of agreement 0.82 to -1.42], and mean bias 0.56 [limits of agreement 1.94 to -0.81], at a plateau pressure of 15 and 30 cmH(2)O, respectively). The ACUTE-19 achieved optimal oxygenation and ventilation before and after acute respiratory distress syndrome induction. CONCLUSIONS: The ACUTE-19 performed accurately in simulated and animal models yielding a comparable performance with a VIVO-50 commercial device. The ACUTE-19 can provide the basis for the development of a future affordable commercial ventilator.

© 2021 Sociedad Española de Anestesiología, Reanimación y Terapéutica del Dolor. Published by Elsevier España, S.L.U. All rights reserved.

Datos de la publicación

ISSN/ISSNe:
0034-9356, 2340-3284

Revista Espanola de Anestesiologia y Reanimacion  ELSEVIER

Tipo:
Article
Páginas:
544-555
PubMed:
36337377
Enlace a otro recurso:
www.scopus.com
Factor de Impacto:
0,258 SCImago ℠
Cuartil:
Q3 SCImago ℠

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Bench study; COVID-19; Mechanical ventilation; Medical capacity surge; SARS-CoV-2; Ventilator performance

Cita

Compartir