Validación preclínica de un respirador de turbina para la ventilación invasiva: el respirador ACUTE-19.
Fecha de publicación:
Fecha Ahead of Print:
Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Casan-Pallardo, M.
- Redondo-Garcia, I, J.
- Viscasillas-Monteagudo, J.
- Gutierrez-Bautista, A.
- Ramirez-Faz, J.
- Alonso-Perez, P.
- Diaz-Lobato, S.
- Neto, A. S.
- Gama de Abreu M
- Pelosi, P.
- Schultz, M. J.
- open source smart breathing system ACUTE-19 investigators
Grupos
Abstract
BACKGROUND: The severe acute respiratory syndrome-coronavirus 2 pandemic pressure on healthcare systems can exhaust ventilator resources, especially where resources are restricted. Our objective was a rapid preclinical evaluation of a newly developed turbine-based ventilator, named the ACUTE-19, for invasive ventilation. METHODS: Validation consisted of (a) testing tidal volume delivery in 11 simulated models, with various resistances and compliances; (b) comparison with a commercial ventilator (VIVO-50) adapting the United Kingdom Medicines and Healthcare products Regulatory Agency-recommendations for rapidly manufactured ventilators; and (c) in vivo testing in a sheep before and after inducing acute respiratory distress syndrome by saline lavage. RESULTS: Differences in tidal volume in the simulated models were marginally different (largest difference 33 ml [95% CI 31 to 36]; P < .001). Plateau pressure was not different (-0.3 cmH(2)O [95% CI -0.9 to 0.3]; P = .409), and positive end-expiratory pressure was marginally different (0.3 cmH(2)O [95% CI 0.2 to 0.3]; P < .001) between the ACUTE-19 and the commercial ventilator. Bland-Altman analyses showed good agreement (mean bias -0.29 [limits of agreement 0.82 to -1.42], and mean bias 0.56 [limits of agreement 1.94 to -0.81], at a plateau pressure of 15 and 30 cmH(2)O, respectively). The ACUTE-19 achieved optimal oxygenation and ventilation before and after acute respiratory distress syndrome induction. CONCLUSIONS: The ACUTE-19 performed accurately in simulated and animal models yielding a comparable performance with a VIVO-50 commercial device. The ACUTE-19 can provide the basis for the development of a future affordable commercial ventilator.
© 2021 Sociedad Española de Anestesiología, Reanimación y Terapéutica del Dolor. Published by Elsevier España, S.L.U. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 0034-9356, 2340-3284
- Tipo:
- Article
- Páginas:
- 544-555
- PubMed:
- 36337377
- Enlace a otro recurso:
- www.scopus.com
- Factor de Impacto:
- 0,258 SCImago ℠
- Cuartil:
- Q3 SCImago ℠
Revista Espanola de Anestesiologia y Reanimacion ELSEVIER
Documentos
- No hay documentos
Filiaciones
Keywords
- Bench study; COVID-19; Mechanical ventilation; Medical capacity surge; SARS-CoV-2; Ventilator performance
Cita
ALONSO JM,MAZZINARI G,Casan M,J,Viscasillas J,Gutierrez A,Ramirez J,Alonso P,Diaz S,Neto AS,DIAZ O,ARGENTE P,Gama de Abreu M,Pelosi P,Schultz MJ,open source smart breathing system ACUTE I. Validación preclínica de un respirador de turbina para la ventilación invasiva: el respirador ACUTE-19. Rev. Esp. Anestesiol. Reanim. 2022. 69. (9):p. 544-555. IF:1,300.
Portal de investigación