In Vivo Verification of Treatment Source Dwell Times in Brachytherapy of Postoperative Endometrial Carcinoma: A Feasibility Study

Autors de IIS La Fe
Autors aliens a IIS La Fe
- Herreros, Antonio
- Barrera-Gomez, Jose
- Abellana, Rosa
- Melo, Joana
- Moutinho, Luis
- Tagliaferri, Luca
- Rovirosa, Angeles
Grups d'Investigació
Abstract
(1) Background: In brachytherapy, there are still many manual procedures that can cause adverse events which can be detected with in vivo dosimetry systems. Plastic scintillator dosimeters (PSD) have interesting properties to achieve this objective such as real-time reading, linearity, repeatability, and small size to fit inside brachytherapy catheters. The purpose of this study was to evaluate the performance of a PSD in postoperative endometrial brachytherapy in terms of source dwell time accuracy. (2) Methods: Measurements were carried out in a PMMA phantom to characterise the PSD. Patient measurements in 121 dwell positions were analysed to obtain the differences between planned and measured dwell times. (3) Results: The repeatability test showed a relative standard deviation below 1% for the measured dwell times. The relative standard deviation of the PSD sensitivity with accumulated absorbed dose was lower than 1.2%. The equipment operated linearly in total counts with respect to absorbed dose and also in count rate versus absorbed dose rate. The mean (standard deviation) of the absolute differences between planned and measured dwell times in patient treatments was 0.0 (0.2) seconds. (4) Conclusions: The PSD system is useful as a quality assurance tool for brachytherapy treatments.
Dades de la publicació
- ISSN/ISSNe:
- 2075-4426, 2075-4426
- Tipus:
- Article
- Pàgines:
- -
- DOI:
- 10.3390/jpm12060911
- PubMed:
- 35743696
- Factor d'Impacte:
- 0,757 SCImago ℠
- Quartil:
- Q2 SCImago ℠
Journal of Personalized Medicine Multidisciplinary Digital Publishing Institute (MDPI)
Documents
- No hi ha documents
Filiacions
Keywords
- brachytherapy; endometrial carcinoma; treatment verification; in vivo dosimetry; plastic scintillator dosimeter
Projectes associats
ESTUDIO OBSERVACIONAL PARA DESCRIBIR RETROSPECTIVAMENTE LA EVOLUCIÓN DE LOS PACIENTES CON CÁNCER DE PRÓSTATA A LO LARGO DE LA ENFERMEDAD A TRAVÉS DE LA REUTILIZACIÓN DE LOS REGISTROS ELECTRÓNICOS DE SALUD CON INTELIGENCIA ARTIFICIAL. ESTUDIO OVERVIEW.
Investigador Principal: ANTONIO CONDE MORENO
JAN-CPR-2018-01 . 2018
CHAIMELEON. ACCELERATING THE LAB TO MARKET TRANSITION OF AI TOOLS FOR CANCER MANAGEMENT.
Investigador Principal: LUIS MARTÍ BONMATÍ
952172 . COMISION EUROPEA . 2020
ESTUDIO DE BRAQUITERAPIA ELECTRÓNICA PARA CARCINOMA CUTÁNEO BASOCELULAR.
Investigador Principal: JOSÉ PÉREZ CALATAYUD
BCCBRACHY