A Monte Carlo study of the relative biological effectiveness in surface brachytherapy.
Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Valdes-Cortez, Christian
- Niatsetski, Yury
Grupos
Abstract
Purpose This work aims to simulate clustered DNA damage from ionizing radiation and estimate the relative biological effectiveness (RBE) for radionuclide (rBT)- and electronic (eBT)-based surface brachytherapy through a hybrid Monte Carlo (MC) approach, using realistic models of the sources and applicators. Methods Damage from ionizing radiation has been studied using the Monte Carlo Damage Simulation algorithm using as input the primary electron fluence simulated using a state-of-the-art MC code, PENELOPE-2018. Two Ir-192 rBT applicators, Valencia and Leipzig, one Co-60 source with a Freiburg Flap applicator (reference source), and two eBT systems, Esteya and INTRABEAM, have been included in this study implementing full realizations of their geometries as disclosed by the manufacturer. The role played by filtration and tube kilovoltage has also been addressed. Results For rBT, an RBE value of about 1.01 has been found for the applicators and phantoms considered. In the case of eBT, RBE values for the Esteya system show an almost constant RBE value of about 1.06 for all depths and materials. For INTRABEAM, variations in the range of 1.12-1.06 are reported depending on phantom composition and depth. Modifications in the Esteya system, filtration, and tube kilovoltage give rise to variations in the same range. Conclusions Current clinical practice does not incorporate biological effects in surface brachytherapy. Therefore, the same absorbed dose is administered to the patients independently on the particularities of the rBT or eBT system considered. The almost constant RBE values reported for rBT support that assumption regardless of the details of the patient geometry, the presence of a flattening filter in the applicator design, or even significant modifications in the photon energy spectra above 300 keV. That is not the case for eBT, where a clear dependence on the eBT system and the characteristics of the patient geometry are reported. A complete study specific for each eBT system, including detailed applicator characteristics (size, shape, filtering, among others) and common anatomical locations, should be performed before adopting an existing RBE value.
© 2022 American Association of Physicists in Medicine.
Datos de la publicación
- ISSN/ISSNe:
- 0094-2405, 2473-4209
- Tipo:
- Article
- Páginas:
- 5576-5588
- DOI:
- 10.1002/mp.15774
- Factor de Impacto:
- 1,174 SCImago ℠
- Cuartil:
- Q1 SCImago ℠
MEDICAL PHYSICS WILEY
Documentos
- No hay documentos
Filiaciones
Keywords
- Monte Carlo; relative biological effectiveness; surface HDR brachytherapy
Proyectos asociados
ESTUDIO OBSERVACIONAL PARA DESCRIBIR RETROSPECTIVAMENTE LA EVOLUCIÓN DE LOS PACIENTES CON CÁNCER DE PRÓSTATA A LO LARGO DE LA ENFERMEDAD A TRAVÉS DE LA REUTILIZACIÓN DE LOS REGISTROS ELECTRÓNICOS DE SALUD CON INTELIGENCIA ARTIFICIAL. ESTUDIO OVERVIEW.
Investigador Principal: ANTONIO CONDE MORENO
JAN-CPR-2018-01 . 2018
CHAIMELEON. ACCELERATING THE LAB TO MARKET TRANSITION OF AI TOOLS FOR CANCER MANAGEMENT.
Investigador Principal: LUIS MARTÍ BONMATÍ
952172 . COMISION EUROPEA . 2020
ESTUDIO DE BRAQUITERAPIA ELECTRÓNICA PARA CARCINOMA CUTÁNEO BASOCELULAR.
Investigador Principal: JOSÉ PÉREZ CALATAYUD
BCCBRACHY
Cita
Valdes C,Niatsetski Y,Perez J,Ballester F,Vijande J. A Monte Carlo study of the relative biological effectiveness in surface brachytherapy. Med Phys. 2022. 49. (8):p. 5576-5588. IF:3,800. (2).