An all-in-one point-of-care testing device for multiplexed detection of respiratory infections.

Fecha de publicación: Fecha Ahead of Print:

Autores de IIS La Fe

  • Ángel Maquieira Catalá

    Autor

  • Sergi Morais Ezquerro

    Autor

Participantes ajenos a IIS La Fe

  • Teixeira, W
  • Pallás-Tamarit, Y
  • Juste-Dolz, A
  • Sena-Torralba, A
  • Gozalbo-Rovira, R
  • Rodríguez-Díaz, J
  • Navarro, D
  • Carrascosa, J
  • Gimenez-Romero, D

Grupos

Abstract

The impact of the COVID-19 pandemic has reinforced the need for rapid, cost-effective, and reliable point-of-care testing (POCT) devices for massive population screening. The co-circulation of SARS-CoV-2 with several seasonal respiratory viruses highlights the need for multiplexed biosensing approaches. Herein, we present a fast and robust all-in-one POCT device for parallel viral antigen and serological analysis. The biosensing approach consists of a functionalized polycarbonate disc-shaped surface with microfluidic structures, where specific bioreagents are immobilized in microarray format, and a portable optoelectronic analyzer. The biosensor quantifies the concentration of viral antigens and specific immunoglobulins G and M for SARS-CoV-2, influenza A/B, adenovirus, and respiratory syncytial virus, using 30 µL of a sample. The semi-automated analysis of 6 samples is performed in 30 min. Validation studies performed with 135 serum samples and 147 nasopharyngeal specimens reveal high diagnostic sensitivity (98-100%) and specificity (84-98%), achieving an excellent agreement (? = 0.937) with commercial immunoassays, which complies with the World Health Organization criteria for POC COVID-19 diagnostic tests. The versatility of the POCT device paves the way for the detection of other pathogens and analytes in the incoming post-pandemic world, integrating specific bioreagents against different variants of concerns and interests.

Copyright © 2022. Published by Elsevier B.V.

Datos de la publicación

ISSN/ISSNe:
0956-5663, 1873-4235

BIOSENSORS & BIOELECTRONICS  ELSEVIER ADVANCED TECHNOLOGY

Tipo:
Article
Páginas:
114454-114454
PubMed:
35696866
Enlace a otro recurso:
www.scopus.com
Factor de Impacto:
2,113 SCImago ℠
Cuartil:
Q1 SCImago ℠

Citas Recibidas en Web of Science: 43

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Adenovirus; COVID-19; Immunoassay; Influenza; Microfluidics; Respiratory syncytial virus; SARS-CoV-2

Compartir