Embedded bleeding detector into a PMMA applicator for electron intraoperative radiotherapy.

Autores de IIS La Fe
Participantes ajenos a IIS La Fe
- Garcia-Gil, Rafael
- Casans, Silvia
- Edith Navarro A
- Garcia-Sanchez, Antonio-Jose
- Rovira-Escutia, Juan J.
- Garcia-Costa, Daniel
- Sanchis-Sanchez, Enrique
- Perez-Calatayud, Ignacio
Grupos
Abstract
PURPOSE: The aim of this work is to present a ready to industrialize low-cost and easy-to-install bleeding detector for use in intraoperative electron radiation therapy (IOERT). The detector works in stand-alone mode and is embedded into a translucent polymethylmethacrylate (PMMA) applicator avoiding any contact with the patient, which represent a novelty compared to previous designs. The use of this detector will prevent dose misadministration during irradiation in the event of accumulation of fluids in the applicator. METHODS: The detector is based on capacitive sensor and wireless power-supply electronics. Both sensor and electronics have been embedded in the applicator, so that any contact with the patient would be avoided. Since access to the tumor can be done through different trajectories, the detector has been calibrated for different tilting angles. RESULTS: The result of the calibration provides us with a fit curve that allows the interpolation of the results at any angle. Comparison of estimated fluid height vs real height gives an error of 1 mm for tilting angles less than 10° and 2 mm for tilting angles greater than 15°. This accuracy is better than the one required by clinic. CONCLUSIONS: The performance of the bleeding detector was evaluated in situ. No interference was observed between the detector and the beam. In addition, a user-friendly mobile application has been developed to help the surgical team making decisions before and during irradiation. The measurement provided by the mobile application was stable during the irradiation process.
Copyright © 2021 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 1120-1797, 1724-191X
- Tipo:
- Article
- Páginas:
- 35-42
- Factor de Impacto:
- 0,747 SCImago ℠
- Cuartil:
- Q2 SCImago ℠
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS ELSEVIER SCI LTD
Citas Recibidas en Web of Science: 1
Documentos
- No hay documentos
Filiaciones
Keywords
- Bleeding detector; Capacitive sensor; Intraoperative radiotherapy
Proyectos asociados
ESTUDIO OBSERVACIONAL PARA DESCRIBIR RETROSPECTIVAMENTE LA EVOLUCIÓN DE LOS PACIENTES CON CÁNCER DE PRÓSTATA A LO LARGO DE LA ENFERMEDAD A TRAVÉS DE LA REUTILIZACIÓN DE LOS REGISTROS ELECTRÓNICOS DE SALUD CON INTELIGENCIA ARTIFICIAL. ESTUDIO OVERVIEW.
Investigador Principal: ANTONIO CONDE MORENO
JAN-CPR-2018-01 . 2018
CHAIMELEON. ACCELERATING THE LAB TO MARKET TRANSITION OF AI TOOLS FOR CANCER MANAGEMENT.
Investigador Principal: LUIS MARTÍ BONMATÍ
952172 . COMISION EUROPEA . 2020
ESTUDIO DE BRAQUITERAPIA ELECTRÓNICA PARA CARCINOMA CUTÁNEO BASOCELULAR.
Investigador Principal: JOSÉ PÉREZ CALATAYUD
BCCBRACHY
Cita
Garcia R,Casans S,Edith A,Garcia A,Rovira JJ,Garcia D,Sanchis E,Perez I,Perez J,Sanchis E. Embedded bleeding detector into a PMMA applicator for electron intraoperative radiotherapy. Phys Med. 2022. 94. p. 35-42. IF:3,400. (2).