Abstract

Capsule endoscopy (CE) is a widely used, minimally invasive alternative to traditional endoscopy that allows visualisation of the entire small intestine. Patient preparation can help to obtain a cleaner intestine and thus better visibility in the resulting videos. However, studies on the most effective preparation method are conflicting due to the absence of objective, automatic cleanliness evaluation methods. In this work, we aim to provide such a method capable of presenting results on an intuitive scale, with a relatively light-weight novel convolutional neural network architecture at its core. We trained our model using 5-fold cross-validation on an extensive data set of over 50,000 image patches, collected from 35 different CE procedures, and compared it with state-of-the-art classification methods. From the patch classification results, we developed a method to automatically estimate pixel-level probabilities and deduce cleanliness evaluation scores through automatically learnt thresholds. We then validated our method in a clinical setting on 30 newly collected CE videos, comparing the resulting scores to those independently assigned by human specialists. We obtained the highest classification accuracy for the proposed method (95.23%), with significantly lower average prediction times than for the second-best method. In the validation of our method, we found acceptable agreement with two human specialists compared to interhuman agreement, showing its validity as an objective evaluation method.

Datos de la publicación

ISSN/ISSNe:
2045-2322, 2045-2322

SCIENTIFIC REPORTS  NATURE PUBLISHING GROUP

Tipo:
Article
Páginas:
17706-17706
PubMed:
33077755
Factor de Impacto:
1,240 SCImago
Cuartil:
Q1 SCImago

Citas Recibidas en Web of Science: 8

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Campos de Estudio

Proyectos y Estudios Clínicos

RED DE INNOVACION EN TECNOLOGIAS MEDICAS Y SANITARIAS - RED ITEMAS

Investigador Principal: JOSÉ VICENTE CASTELL RIPOLL

RD09/0077/00156 . INSTITUTO DE SALUD CARLOS III . 2010

WIBEC. WIRELESS IN-BODY ENVIRONMENT COMMUNICATIONS.

Investigador Principal: VICENTE PONS BELTRÁN

675353 . COMISION EUROPEA . 2016

Optimización del dispositivo Endoworm de asistencia para la realización de enteroscopia.

Investigador Principal: VICENTE PONS BELTRÁN

PI18/01365 . INSTITUTO DE SALUD CARLOS III . 2019

ESTUDIO DE FASE III, ALEATORIZADO, A DOBLE CIEGO, CONTROLADO CON PLACEBO SOBRE TERAPIA ADYUVANTE CON VEMURAFENIB (RO5185426) EN PACIENTES CON MELANOMA CUTÁNEO CON MUTACIÓN DE BRAF RESECADO QUIRÚRGICAMENTE CON ALTO RIESGO DE RECIDIVA.

Investigador Principal: ROBERTO PEDRO DÍAZ BEVERIDGE

GO27826 . 2013

ESTUDIO ALEATORIZADO Y ABIERTO, PARA COMPARAR DURANTE 26 SEMANAS EL EFECTO DEL TRATAMIENTO CON GEL INTESTINAL DE LEVODOPA-CARBIDOPA (GILC) Y DEL TRATAMIENTO MÉDICO ÓPTIMO (TMO) SOBRE LOS SÍNTOMAS NO MOTORES (SNM) EN PACIENTES CON ENFERMEDAD DE PARKINSON AVANZADA - ESTUDIO INSIGHTS.

Investigador Principal: IRENE MARTÍNEZ TORRES

M12-927 . 2015

A RANDOMIZED, DOUBLE-BLIND, PLACEBO-CONTROLLED PHASE 4 STUDY TO EVALUATE THE EFFICACY AND SAFETY OF ENTYVIO (VEDOLIZUMAB IV) IN THE TREATMENT OF CHRONIC POUCHITIS./ ESTUDIO DE FASE IV ALEATORIZADO, DOBLE CIEGO Y CONTROLADO CON PLACEBO PARA EVALUAR LA EFICACIA Y LA SEGURIDAD DE ENTYVIO (VEDOLIZUMAB I.V.) EN EL TRATAMIENTO DE LA POUCHITIS CRÓNICA.

Investigador Principal: GUILLERMO BASTIDA PAZ

VEDOLIZUMAB-4004 . 2017

ESTUDIO DE INDUCCIÓN MULTICÉNTRICO, ALEATORIZADO, DOBLE CIEGO, CONTROLADO CON PLACEBO PARA EVALUAR LA EFICACIA Y SEGURIDAD DE UPADACITINIB (ABT 494) EN PACIENTES CON COLITIS ULCEROSA DE MODERADA A GRAVE.

Investigador Principal: MARISA IBORRA COLOMINO

M14-675 . 2019

ESTUDIO OBSERVACIONAL PARA LA EVALUACIÓN DE DIFERENTES PREPARACIONES PARA LA REALIZACIÓN DE CÁPSULA ENDOSCÓPICA MEDIANTE UN SISTEMA DE DETECCIÓN AUTOMÁTICA.

Investigador Principal: VICENTE PONS BELTRÁN

VPB-MAC-2019-01 . 2020

ESTUDIO EN FASE III, CIEGO, ALEATORIZADO, MULTICENTRICO, PARA LA EVALUACION DE SEGURIDAD Y EFICACIA DE CITRAFLEET (R) (PICOSULFATO SODICO) COMO TRATAMIENTO EVACUANTE PREVIO A EXPLORACION ENDOSCOPICA COLONICA.

Investigador Principal: VICENTE PONS BELTRÁN

CF-PICOS2006

IMPACTO DE LA PAPILOTOMIA DURANTE LA COLANGIOPANCREATOGRAFÍA RETRÓGRADA ENDOSCÓPICA (CPRE) PARA LA IMPLANTACION DE UN STENT BILIAR METALICO NO CUBIERTOEN PACIENTES CON ICTERICIA OBSTRUCTIVA POR TUMORES MALIGNOS DE PANCREAS

Investigador Principal: VICENTE PONS BELTRÁN

PAP-ADE-01

ESTUDIO ALEATORIZADO, ENMASCARADO, CONTROLADO CON PLACEBO SOBRE LOS EFECTOS DE ADALIMUMAB INTRALESIONAL EN ESTENOSIS INTESTINALES DE PACIENTES CON ENFERMEDAD DE CROHN.

Investigador Principal: LIDIA ARGÜELLO VIUDEZ

CSAI

Cita

Compartir