Analysis of the 'Endoworm' prototype's ability to grip the bowel in in vitro and ex vivo models.

Fecha de publicación: Fecha Ahead of Print:

Autores de IIS La Fe

Participantes ajenos a IIS La Fe

  • Tobella J
  • Santonja A
  • Sánchez C
  • Campillo-Fernández AJ
  • Vidaurre A

Grupos

Abstract

Access to the small bowel by means of an enteroscope is difficult, even using current devices such as single-balloon or double-balloon enteroscopes. Exploration time and patient discomfort are the main drawbacks. The prototype 'Endoworm' analysed in this paper is based on a pneumatic translation system that, gripping the bowel, enables the endoscope to move forward while the bowel slides back over its most proximal part. The grip capacity is related to the pressure inside the balloon, which depends on the insufflate volume of air. Different materials were used as in vitro and ex vivo models: rigid polymethyl methacrylate, flexible silicone, polyester urethane and ex vivo pig small bowel. On measuring the pressure-volume relationship, we found that it depended on the elastic properties of the lumen and that the frictional force depended on the air pressure inside the balloons and the lumen's elastic properties. In the presence of a lubricant, the grip on the simulated intestinal lumens was drastically reduced, as was the influence of the lumen's properties. This paper focuses on the Endoworm's ability to grip the bowel, which is crucial to achieving effective endoscope forward advance and bowel folding.

Datos de la publicación

ISSN/ISSNe:
0954-4119, 2041-3033

PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE  SAGE PUBLICATIONS LTD

Tipo:
Article
Páginas:
468-477
PubMed:
31984867
Factor de Impacto:
0,389 SCImago
Cuartil:
Q2 SCImago

Citas Recibidas en Web of Science: 4

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Enteroscopy, grip force measurement, medical control systems, small bowel

Cita

Compartir