Guide for quality control and safety in clinical linear accelerators

Fecha de publicación: Fecha Ahead of Print:

Autores de IIS La Fe

Participantes ajenos a IIS La Fe

  • Almendral Manzano, P.
  • Ambroa Rey, E.
  • Ferrer, C.
  • Martinez-Albaladejo, M.
  • Pérez-Alija Fernández, J.
  • Tortosa Oliver, R.
  • Vilches Pacheco, M.

Grupos

Abstract

The main objective of this document has been to establish a guide for the design of the quality control programme in linear accelerators with isocentric assembly, incorporating the approaches of statistical process control and risk analysis. Throughout the text, it is emphasized that the guide is not prescriptive, and adaptation to the specific needs of each centre is encouraged. A brief description of the recommended tests is presented. A set of fundamental parameters and complementary ones are suggested, taking into account the results obtained from a survey of SEFM members, as well as proposals from other scientific societies, manufacturer comments, and consensus from this working group. Frequency and tolerance recommendations are provided for representative accelerators as a reference. This document presents a summarized version of the guide, which is accessible in Revista de Física Médica or on the website of the SEFM, providing a synthesis of the main recommendations. © 2026 Associazione Italiana di Fisica Medica e Sanitaria.

Datos de la publicación

ISSN/ISSNe:
1120-1797, 1724-191X

Physica Medica-European Journal of Medical Physics  ELSEVIER SCI LTD

Tipo:
Article
Páginas:
-
PubMed:
41679132
Enlace a otro recurso:
www.scopus.com
Factor de Impacto:
0,747 SCImago ℠
Cuartil:
Q2 SCImago ℠

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • accuracy; Article; body position; documentation; dosimetry; practice guideline; process design; quality control; radiation dose; radiation field; radiation tolerance; radiotherapy; risk assessment; risk management; safety

Compartir